9
Checking the meter
4
Apply the control solution to the
centre of the test area at the tip
of the cassette using the tip of
the brush.
H
Do not hold the meter with the
test area pointing upwards
while you apply control solution
because the solution could
enter the meter and soil it.
96
5
Take the brush away from the
test area as soon as Test in
progress appears on the screen.
The test area has absorbed
sufficient control solution, the
beep tone sounds and the control
result is displayed.
6
Press .
The test ends after 5 seconds
and the meter displays the result.