Appendix E

SCXI-1340 Installation

Follow these steps to install the SCXI-1340:
1. Make sure that the computer and the SCXI chassis are turned off.
2. Install the SCXI module in the chassis.
3. Plug the mounting bracket connector onto the module rear signal connector (see Figure E-1).
An alignment tab on the bracket enters the upper board guide of the chassis.
4. Screw the mounting bracket to the threaded strips in the rear of the chassis.
5. Connect the loose end of the cable assembly to the MIO-16 or DIO-24 board rear signal
connector.
Check the installation.
After step 1, the order of these steps is not critical; however, it is easier to locate the correct
position for the mounting bracket with a module installed in the chassis. If you are attaching a
cable to the breakout connector, installation is easiest if you attach the second cable before
installing the SCXI-1340.
© National Instruments Corporation
Mounting Bracket
Connector
Step 3
SCXI-1163 Rear
Signal Connector
Mounting Bracket

Figure E-1. SCXI-1340 Installation

E-3
SCXI-1163 Cabling
Rear Panel
Step 4
Step 5
50-Pin Female
Connector to
MIO-16 or
DIO-24 Board
Male Breakout
Step 4
Connector
SCXI-1163 User Manual