Automatic Calibration - Pace SODRTEK ST 325 Operation And Maintenance Manual

Digital convective soldering/desoldering system
Table of Contents

Advertisement

7. Use the Scroll Down Button ( ) to select the desired temperature scale.
8. Press the Select Button or the Scroll Up Button (
of the Auto Vac, (e.g., "Auto Vac = On").
9. Use the Scroll Down Button ( ) to select the desired Auto Vac state.
10. Press the Select Button or the Scroll Up Button (
Profiles?". If you want to delete a profile, press the Scroll Up Button (
delete a profile, press the Scroll Down Button ( ). The LCD will now display "Exit Setup?"
11. Use the Scroll Up Button (
the program will cycle to the beginning and the LCD will display "Password Needed?".
12. If the Scroll Up Button (
13. Use the Scroll Keys (
Key. The Screen will now ask you to confirm the deletion, (e.g., "24 are you sure?").
14. Use the Scroll Up Button (
15. If no is selected, the system will prompt the user to exit the set up mode. Use the Scroll Up
Button (
) for yes and the Scroll Down Button ( ) for no. If no is selected, the program will
cycle to the beginning and the LCD will display "Password Needed?".

Automatic Calibration

The ST 325 System provides precision control of temperature thanks to the closed-loop controlled circuit
design. The temperature sensor is located in the heater but utilizing the PACE Thermocouple Nozzle can
yield more accurate results due to moving the temperature sensor closer to the component. The
Automatic Calibration Procedure allows the system to calibrate for various style components as well as to
adjust for any minor offsets in temperature due to heater differences and blower speeds. Doing this will
ensure the most accurate temperature readings.
NOTE: For best results, hold a PCB up to the Thermocouple Nozzle at the same distance as the
nozzle would be from the PCB during the removal or installation process. During the
calibration it is important to avoid holding the PCB in one position too long; this will avoid
board damage during the Automatic Calibration Procedure.
Entering the Automatic Calibration Mode
1. Turn the Power Switch Off.
2. Install the Thermocouple Nozzle.
3. Press and hold the Scroll Up Button (
Release these buttons when software development date appears, e.g., "Dev Date 8-20-04."
©2013 PACE Inc., Southern Pines, NC
All Rights Reserved
) for yes and the Scroll Down Button ( ) for no. If no is selected,
) is selected, the LCD will flash "Delete Profile and Number -- ?"
) to select the desired profile number (1-40) and then press the Select
) for yes and the Scroll Down Button ( ) for no.
) and Select Button while turning the Power Switch On.
) to save. The LCD will now display the status
) to save. The LCD will now display "Delete
). If you do not want to