Default Factory Settings; Available Tips; System Calibration - Pace WJS 100 Operation And Maintenance Manual

Soldering system
Table of Contents

Advertisement

a) Press and release the Scroll Up
operation.
b) Press and release the Scroll Down
procedure. Go back to step 4.

Default Factory Settings

The WJS-100 system comes equipped with a number of features, which may be adjusted by the user.
Listed below are the features and default settings of each. To change and/or learn about any of
these features, refer to the applicable part of the "Customizing Your System" section of this manual.
Feature
Password
Default Temperature Scale (°C/°F)
"HI" (Upper) Temperature Limit
"LO" (Lower) Temperature Limit
Set Temperature
Tip Offset Constant
Temperature Setback
Auto Off
Available Tips – ONLY USE 1128 SERIES TIPS with GOLD END CONNECTORS
1128-0001-P1 TWJS High Performance TIP, 1/32" Conical Sharp Extended
1128-0002-P1 WJS High Performance TIP 1/64" Conical Sharp
1128-0003-P1 WJS High Performance TIP 1/64" Conical, Sharp, Bent 30 Degree
1128-0008-P1 WJS High Performance TIP 3/64" 30 Degree Chisel
1128-0010-P1 WJS High Performance TIP 13/64" Chisel
1128-0012-P1 WJS High Performance TIP 1/32" 30 Degree Chisel
1128-0013-P1 WJS High Performance TIP 3/32" 30 Degree Chisel
1128-0019-P1 WJS High Performance TIP 1/16" 30 Degree Chisel
1128-0032-P1 WJS High Performance TIP MINIWAVE
1128-0037-P1 WJS High Performance TIP 0.25" Knife Blade

System Calibration

All systems are tested for temperature accuracy at the factory and can be checked for
calibration according to requirements. No internal adjustments can be made to the power
supply.
To Adjust Digital Temperature Control Systems
1. After the system/tip has stabilized, measure the tip temperature with whatever method
you use.
2. Turn off power source and turn it back on while holding down the program (round) key
and the up key.
3. The system will prompt you to enter the measured temperature using the keypad.
4. Press the program key and the system will restart.
5. To clear the calibration, remove the handpiece while the power source is ON.
©2010 PACE Inc., Annapolis Junction, Maryland All Rights Reserved
Key to exit Set-Up Mode and return to normal
Key to return to the start of the Set-Up Mode
Factory Setting
None Entered
°F for 115 VAC Systems
°C for 230 VAC Systems
454 °C (850 °F)
204 °C (400 °F)
"OFF"
"0"
Enabled, 30 minutes
Enabled, 30 minutes