Auto Off Mode; Using The Tw 100 Calibration Procedure - Pace HW 100 HeatWise Operation And Maintenance Manual

Soldering systems, voltage 115/230 vac
Hide thumbs Also See for HW 100 HeatWise:
Table of Contents

Advertisement

2. MANUAL ACTIVATION: The operator can manually force the system into Temperature
Setback by performing the following procedure.
a. Press and hold the Scroll Down
b. Release both keys.
Exiting Temperature Setback: Listed below are 3 ways to exit Temperature Setback.
1. Press and release any key on the front panel ((
method.
2. Wipe the hot handpiece tip on a wet sponge to thermally load the tip.
3. Turn the Power Switch "OFF" ("0") and then back "ON" ("l").
Set Tip Temperature values will be restored. For optimum performance, do not attempt to use the
attached handpiece until the Set Tip Temperature is achieved and the LED indicator is green.
Auto Off Safety System Mode
When enabled, the Auto Off safety system of the TW 100 system turns off the power to the handpiece 10-
90 minutes after entering Temperature Setback. When the system has entered Temperature Setback, an
Auto Off timer within the system circuitry will start running if Auto Off is turned on. Refer to number 10 in
the Programming Menu. When Auto Off has activated, the LED Display will blink "OFF".
1. If any key is pressed during the selected time out period, the Auto Off timer is reset. The
system will return to normal operation.
2. At the end of the time out period, the system will enter Auto Off. Power is turned off to the
heater and the LED Display will show a flashing "OFF " and the LED indicator will turn red.
Exiting Auto Off: Auto Off can be exited; returning to normal operation by:
1. Pressing and releasing any key on the front panel. (
2. By turning the Power Switch OFF ("0") and then back ON ("l").
Using the TW 100 Calibration Feature
The new TW 100 systems feature an all-new calibration procedure. The system is calibrated so the
display matches the measured tip temperature using your own testing method. This new PACE feature
allows any type of temperature verification device to be utilized to complete the simple procedure.
Calibration can be completed, as often or as little as you like, in order to meet your own internal
specifications.
The steps for the procedure are:
1. Remove any offset from the system by disconnecting the handpiece from the system; the
LED should go to amber.
2. Set the TW 100 to 700°.
3. Record the actual temperature of the tip from your temperature verification device.
4. Place the TW 100 into the Calibration mode. This can be achieved by starting with the
system power switch in the off position. Press and hold the Program key and the UP
while turning the system on. Release both keys when the display reads "1-7", once released
the display will read "tIP." Please note that if the system has already been calibrated that the
display will read "CSO" at this time. To clear this, simply momentarily remove the handpiece
from the system.
5. Use the UP
temperature verification device.
6. Press the PROGRAM key to exit. Upon exiting, LED will illuminate green.
NOTE: The system will display "CSO" (Clear System Offset) if an offset exists within the system.
©2004 PACE Inc., Annapolis Junction, Maryland
All Rights Reserved
Key and the Scroll Up
and DOWN
keys to enter the temperature that you recorded by the
Key.
),
or
). This is the preferred
,
or
), or
key
Page 10 of 17