Performing the Calibration Procedure

NOTE:
The calibration procedure can only be performed from the front panel or using the SCPI
language commands. If your unit is set to the COMPatibility programming language, use
either the front panel
programming language to SCPI.
Table B-1 lists the equipment required for calibration. Figure B-1 shows the test setup.
You do not have to do a complete calibration each time. If appropriate, you may calibrate only the
voltage or current and proceed to "Saving the Calibration Constants". However, the voltage or current
calibration sequence must be performed in its entirety. The following parameters may be calibrated:
voltage programming and measurement
overvoltage protection (OVP)
current programming and measurement
low range measurement
ac current measurement
Table B-2. Verification Test Record for HP 66311A
Report No.____________ Date_____________
Specification
Voltage Programming and Measurement
V
V
Current Programming and Measurement (High Range)
I
Current Measurement (Low Range)
I
I
out
Current Sink Measurement
I
sink
I
sink
Address
Minimum
Recorded Results
−10 mV
_____mV
−5 mV
_____mV
out
14.982 V
_______V
−9.5 mV
_______V
out
−1.33 mA
_____mA
2.9972 A
_______A
−6.5 mA
_______A
out
−2.5 µA
_____µA
out
−22.5 µA
_____mA
−22.5 µA
_____mA
−5.1 mA
_____A
key or the SYSTem:LANGuage command to change the
Verification and Calibration - B
Maximum
Specification
+10 mV
V
+5 mV
out
15.018 V
V
+9.5 mV
out
1.33 mA
3.0028 A
I
+6.5 mA
out
+2.5 µA
I
out
+22.5 µA
I
out
+22.5 µA
I
sink
I
+5.1 mA
sink
127