Current Output Checkout - Agilent Technologies E3647A User Manual

Table of Contents

Advertisement

Chapter 1 Quick Start
Output Checkout

Current Output Checkout

The following steps check basic current functions with a short across the power
supply's output.
Power
1 Turn on the power supply.
Make sure that the output is disabled. The
2 Connect a short across (+) and (-) output terminals with an insulated test lead.
Use a wire size sufficient to handle the maximum current (See "Table 2-1 Wire
Rating" on page 34 in chapter 2).
Output
On/Off
3 Enable the output.
The
CV
that the display is in the meter mode.
Display
Limit
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the
voltage limit to 1.0 volt to assure CC operation. The
go back to normal mode, press the
several seconds.
Voltage
Current
5 Set the knob to the current control to check that the front-panel ammeter
properly responds to knob control.
Turn the knob clockwise or counter clockwise when the display is in the meter mode
(the
Limit
the voltmeter indicates nearly zero (the voltmeter will show the voltage drop caused
by the test lead). The flashing digit can be adjusted by turning the knob.
1
6 Ensure that the current can be adjusted from zero to the full rated value.
7 Turn off the power supply and remove the short from the output terminals.
1
You can use the resolution selection keys to move the flashing digit to the right or
left when setting the current.
or
annunciator turns on depending on the resistance of the test lead. Notice
CC
annunciator is off). Check that the ammeter responds to knob control and
annunciator is on
OFF
annunciator will be flashing). Adjust the
Limit
annunciator will turn on. To
CC
Display
key again or let the display time out after
Limit
1
21