Fluke 5700A Series II Operator's Manual page 169

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

CAL_CONF
Description
Sequential command.
Sets the calibration specifications to a confidence level of either 99% or 95%.
CONF95 or CONF99: Sets 95% or 99% specifications.
Parameters:
CAL_CONF?
Description
Sequential command.
Returns the current calibration confidence level.
None
Parameters:
Response
CONF95 (indicates 95% specifications)a or CONF99 (indicates 99% specifications)
CAL_CONST?
Description
Sequential command.
Returns the value of a particular calibration constant from a particular group of constants.
1. CAL
Parameters:
CHECK (From calibration check constants)
PREV
2. Symbolic name of the calibration constant desired (see Appendix D).
Response
(Float)
Example:
CAL_CONST? CHECK, KV6
Could return: 6.5000010E+00
CAL_DATE?
Description
Sequential command.
Returns the date of the most recent calibration of the specified type.
B5725
Parameter
CAL
WBFLAT (Last wideband flatness calibration)
WBGAIN (Last wideband gain calibration)
ZERO
Response
(Integer) Date as MMDDYY, DDMMYY, or YYMMDD depending on the setting of
DATEFMT.
CAL_DAYS?
Description
Sequential command.
Returns the number of days elapsed since the last calibration activity of the specified type.
B5725
Parameter
CAL
WBFLAT (Last wideband flatness calibration)
WBGAIN (Last wideband gain calibration)
ZERO
Response
(Integer) Number of elapsed days
Table 5-2. Commands (continued)
(From active calibration constants)
(From previous calibration constants)
The value of the constant
(Last 5725A Amplifier calibration)
(Last calibration of the calibrator)
(Last dc V zero calibration)
(Last 5725A Amplifier calibration)
(Last calibration of the calibrator)
(Last dc V zero calibration)
Remote Operation
5
Commands
5-19