Fluke 5700A Series II Operator's Manual page 262

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II
Operators Manual
6. Reverse the dual-banana connector so that the center connector is connected to LO;
7. To store the new constants, set the CALIBRATION switch to ENABLE and press
8. You have now completed the wideband gain calibration. Set the CALIBRATION
7-20
then press E. The display shows the following:
PREV
MENU
After a short time, a message appears indicating that the wideband calibration is
complete.
"Store Values." To discard the constants, press "DONE with Cal" and answer "YES"
when the display asks for verification.
switch to NORMAL, disconnect the wideband cable, and press r.