Fluke 5700A Series II Operator's Manual page 258

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II
Operators Manual
8. Press O to activate the calibrator output. Then turn the output adjustment knob on
9. Se the rear panel CALIBRATION switch to ENABLE. Press E on the front
10. The range calibration is now complete. Set the rear panel CALIBRATION switch to
7-16
the calibrator until you achieve a null on the 845A Null Detector.
panel. The calibrator will now calculate a new range constant multiplier for the 220V
dc range, and will store it in non-volatile memory.
the NORMAL position, disconnect the external standards, and press R to reset the
calibrator to its newly calibrated ranges.