The adjustment for calibrating the meter requires disassembling the meter.
Refer to the diagrams and access procedures in the 70 Series Service
Manual.
1. Verify that the calibrator is set to 0V dc in standby. Press r if it is not.
2. Turn on the meter, and set its function switch to dcV.
3. Connect a set of test leads to the meter as shown in Figure 4-12.
4. Set the calibrator to 3V dc and press O .
5. Adjust R8 (located just to the left of the rotary switch on the pca) in the meter for a
display reading of +3.00V dc ±.001V.
6. Press r on the calibrator. Disconnect and turn off the meter. Calibration is
complete.