5700A/5720A Series II
Operators Manual
BTYPE?
Description
Parameter
Response
Example:
CAL_ADJ
Description
Parameter
CAL_CHK
Description
Parameter
CAL_CLST?
Description
Parameter
Response
Example:
5-18
Table 5-2. Commands (continued)
Sequential command.
Returns the model numbers of auxiliary amplifiers selected for voltage and current boost.
None
(String, string) VB<model number>,IB<model number>
BTYPE?
Returns: "VB5725,IB5725" if Model 5725A is selected for voltage boost and Model 5725A is
selected for current boost.
Overlapped long-term command, ignored if not in remote.
Performs the internal calibration. This command should be sent after the last CAL_REF
command in the sequence of calibration. (See CAL_REF for an example.)
None
Overlapped long-term command, ignored if not in remote.
Starts calibration check. Results are available via the CAL_SLST? or CAL_SHIFT?
commands.
None
Sequential command.
Returns a list of names values of a particular group of calibration constants.
CAL
(Returns active calibration constants)
CHECK (Returns calibration check constants)
PREV
(Returns previous calibration constants)
(String)
"<EOL>
<total number of constants> <EOL>
<name> <value> <EOL>
<name> <value> <EOL>
(cont)
"
CAL_CLST?
Could return:
"
424
D3P, 3.9817876E+02
(Continues for about 400 pairs of names and values.)
"