2-7. Accessing the Fuse

To prevent instrument damage, verify that the correct fuse is
installed for the line voltage setting.
The line power fuse is accessible on the rear panel. The fuse rating label to the right of
the fuse holder (labeled F1) shows the correct replacement fuse rating for each line
voltage setting. To check or replace the fuse, refer to Figure 2-1 and proceed as follows:
1. Disconnect line power.
2. Using a standard screwdriver, loosen the fuse holder by turning the slot labeled F1
until the cap and fuse pop free.
3. Replace the fuse and holder.
CHASSIS
GROUND
S2
S3
S4
Caution
VOLTAGE
VOLTAGE
FUSE-F1
FUSE-F1
SELECTION
SELECTION
S2 S3 S4
S2 S3 S4
100V
200V
T 3A
110V
220V
250V
115V
230V
(SB)
120V
240V
!
FUSE -F1
Figure 2-1. Accessing the Fuse
T 1.5A
250V
(SB)
Installation
2

Accessing the Fuse