5-45. Remote Program Examples
The following programs are written in BASIC for the Fluke 1722A Instrument
Controller.
5-46. Printing Main Output Calibration Shift Results
This program generates a printout of the changes in the magnitude of self correction
generated at the last calibration. The printout contains the following information:
•
A heading that shows the date, time, and calibrator's serial number
•
Modules present (installed or attached)
•
The most recent calibration dates and temperatures
•
Data about the external standards used
•
Values of the externally-calibrated internal references
•
DC voltage output shifts for each range
•
AC voltage output shifts for each range
•
AC and dc current output shifts for each range
•
Resistance shifts for each value
•
5725 Boost Amplifier output shifts (if attached)
•
Wideband AC Module (Option 5700-03) output shifts (if installed)
1Ø
REM THIS PROGRAM GENERATES A PRINTOUT OF
2Ø
REM MAIN OUTPUT SHIFT RESULTS
3Ø
INIT PORT Ø
4Ø
CLEAR PORT Ø
5Ø
PRINT @4, "CAL_PR CAL"
6Ø
END
! INITIALIZE THE INTERFACE
! SET THE 57ØØA TO ITS POWER-UP STATE
! START CAL REPORT PRINTING
Remote Operation
Remote Program Examples
5
5-61