Fluke 5700A Series II Operator's Manual page 34

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II
Operators Manual
Range
99 % Confidence Level
220 mV
2.2 V
11 V
22 V
220 V
1100 V
95 % Confidence Level
220 mV
2.2 V
11 V
22 V
220 V
1100 V
Notes:
DC Zeros calibration required every 30 days.
For fields strengths >1 V/m but ≤ 3 V/m, add 0.01 % of range.
1.
1-16
5700A Series II DC Voltage Specifications
± 5 °C from calibration temperature
Resolution
24 Hours
10 nV
6.5 + .75
100 nV
3.5 + 1.2
1 µV
3.5 + 3
1 µV
3.5 + 6
10 µV
5 + 100
100 µV
7 + 600
10 nV
5.5 + 0.6
100 nV
3.5 + 1
1 mV
3 + 3.5
1 mV
3 + 6.5
10 mV
4 + 80
100 mV
6 + 500
Absolute Uncertainty
90 Days
180 Days
1 Year
± (ppm output + µV)
7 + .75
8 + .75
9 + .8
6 + 1.2
7 + 1.2
8 + 1.2
5 + 4
7 + 4
5 + 8
7 + 8
6 + 100
8 + 100
9 + 100
8 + 600
10 + 600
11 + 600
6 + 0.6
7 + 0.6
8 + 0.6
5 + 1
6 + 1
4 + 3.5
6 + 3.5
7 + 3.5
4 + 6.5
6 + 6.5
7 + 6.5
5 + 80
7 + 80
8 + 80
7 + 500
8 + 500
9 + 500
Relative Uncertainty
[1]
± 1 °C
24 Hours
90 Days
2.5 + .5
2.5 + 1.2
4 + 1.2
8 + 4
1.5 + 3
3.5 + 4
8 + 8
1.5 + 6
3.5 + 8
2.5 + 100
4 + 100
3 + 600
4.5 + 600
2 + 0.4
3.5 + 0.4
7 + 1
2 + 1
1.2 + 3
1.2 + 6
2 + 80
3.5 + 80
2.4 + 500
4 + 500
4 + .5
3.5 + 1
3 + 3.5
3 + 7