Fluke 5700A Series II Operator's Manual page 186

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5700A/5720A Series II
Operators Manual
*PUD?
Description
Parameter
Response
Example
RANGE?
Description
Parameter
Response
RANGELCK
Description
Parameter
RCOMP
Description
Parameter
REFOUT?
Description
Parameter
Response
5-36
Table 5-2. Commands (continued)
Sequential command.
Returns the contents of the *PUD (Protected User Data) memory.
None
#(non-zero digit) (digits) (user data)
The non-zero digit specifies the number of characters that will follow in the \<digits\> field.
These characters are 0 through 9 (ASCII 48 through 57 decimal). The value of the number
in the \<digits\> field defines the number of user data bytes that follow in the \<user data\>
field. The maximum response is 64 characters.
*PUD?
Returns: "205FLUKE" assuming that this is stored as in the example for PUD* above.
Sequential command.
Returns the present output range.
None
(String) Symbolic name of the range from Table 5-4.
Overlapped command, ignored if not in remote.
Locks or unlocks the present output range. The range automatically unlocks if the output
function changes, for example from dc volts to dc current.
ON
(Locks the range)
OFF
(Unlocks the range)
Overlapped command, ignored if not in remote.
While a resistance output 19 kΩ or lower is selected, RCOMP activates or deactivates two-
wire ohms compensation circuitry.
ON
(Turns on the two-wire compensation circuitry)
OFF
(Turns off the two-wire compensation circuitry)
Sequential command.
Returns the value of the reference, which is the output value of the calibrator the last time a
new reference was established with an OUT, NEW_REF, or MULT.
None
1. (Float) Output amplitude
2. (String) Units (V, A, OHM, or DBM)
3. (String) Frequency (0 if dc)