Fluke 5700A Series II Operator's Manual page 59

Multi-function calibrator
Hide thumbs Also See for 5700A Series II:
Table of Contents

Advertisement

5725A Amplifier:
11 A
40 - 1 k
1 k - 5 k
5 k - 10 k
Notes:
Maximum output from 5720A terminals is 2.2 A. Uncertainty specifications for 220 µA and 2.2 mA ranges are increased by a factor of 1.3, plus
2 µA when supplied through 5725A terminals. Specifications are otherwise identical for all output locations.
1.
Stability specifications are included in the Absolute Uncertainty values for the primary specifications.
Temperature coefficient is an adder to uncertainty specifications that does not apply unless operating more than ±5 °C from calibration
2.
temperature.
3.
For larger resistive loads multiply uncertainty specifications by: (
4.
1.5 V compliance limit above 1 A. 5725A Amplifier may be used in range-lock mode down to 1 A.
5.
For resistive loads within rated compliance voltage limits.
6.
For outputs from the Aux Current terminals, the maximum resistive load for full accuracy is 1 kΩ. For larger resistive loads, multiply the
uncertainty as described in Note 3.
Minimum Output ....................................................9 µA for 220 µA range, 10 % on all other ranges. 1 A minimum for 5725A.
Inductive Load Limits............................................400 µH (5700A/5720A, or 5725A). 20 µH for 5700A/5720A output >1 A.
Power Factors........................................................5700A/5720A, 0.9 to 1; 5725A, 0.1 to 1. Subject to compliance voltage
Frequency:
Range (Hz)..........................................................10.000 - 11.999, 12.00 - 119.99, 120.0 - 1199.9, 1.200 k - 10.000 k
Uncertainty ....................................................±0.01 %
Resolution .....................................................11,999 counts
Settling Time ......................................................5 seconds for 5700A/5720A ranges; 6 seconds for 5725A 11 A range; +1
Overshoot...........................................................<10 %
75 + 100
20 + 75
100 + 150
40 + 75
200 + 300
100 + 75
limits.
second for amplitude or frequency range change.
Introduction and Specifications
30 + 75
3
50 + 75
100 + 75
2
actual
load
)
maximum
load
for
full
accuracy
Electrical Specifications
± (% output)
[5]
3
0.05
[5]
0.12
[5]
0.5
1
1-41