Any protocols for which multiple versions are available have a right-pointing arrowhead (>)
listed next to their name.
3 Double-press the arrowhead for the protocol of interest.
The Select Protocol's Default Version dialog box opens.
4 Press the version that you want to make the default, then press Select.
You are returned to the Protocols screen and the update to the default protocol version has
been applied.

Install firmware updates

When Agilent releases a new version of the firmware software (i.e., the software that runs on the
system's touchscreen) users with an Advanced user account can install the new firmware
through a USB disk or through an online connection.
Upload firmware from a USB disk
1 Save the zip folder containing new firmware files to a USB disk. (Do not use an encrypted USB
disk.)
2 Insert the USB disk into an available USB port on the front of the instrument.
3 From the Home screen, press Settings.
The
4 Press System Settings.
The
5 Press Firmware Update.
The
6 Select USB Disk.
The screen displays a browser of the USB disk contents.
7 In the browser, navigate to the zip folder containing the new firmware files.
8 Press the zip folder to select it.
Figure 30 Firmware Update – USB Disk selected
9 Press Update.
Magnis and MagnisDx NGS Prep System User Guide
Settings screen
opens.
System Settings screen
Firmware Update screen
opens.
opens.
48