Agilent Technologies Magnis G9710A User Manual page 47

Ngs prep system, for in vitro diagnostic use, for research use only, not for use in diagnostic procedures
Table of Contents

Advertisement

11 Press Close.
You are returned to the Protocols screen. The new protocol is now listed.
Download a protocol from the Agilent website
1 From the Home screen, press Settings.
The
2 Press Protocol Update.
The
3 Press Update Protocol.
The
4 At the top of the screen, select Online Download.
The screen displays the protocols available for online download.
5 Mark the protocol(s) that you want to download and install.
Figure 29 Protocol Update screen with protocol file selected online
6 Press Download.
A message box opens notifying you that the protocol was downloaded.
7 Press OK to close the message box.
You are returned to the Protocol Update screen. The protocol has been installed.
8 Press Close.
You are returned to the Protocols screen. The new protocol is now listed.
Change the default version for a protocol
When setting up a protocol run, the system uses the default version of the selected protocol.
1 From the Home screen, press Settings.
The
2 Press Protocol Update.
The
Magnis and MagnisDx NGS Prep System User Guide
Settings screen
opens.
Protocols screen
opens
Protocol Update screen
Settings screen
opens.
Protocols screen
opens, listing all protocols on the system.
opens.
47