Figure 21 Diagnostic Test Report screen
7 Review the report. Make note of any test items marked as Failed.
If any of the items failed the diagnostic test, you will see an error icon near the bottom of the
screen, like the one shown below. Press directly on the icon to view more information about
the failed test items.
8 When finished, press Close to close the report.

View reports for diagnostic tests and instrument health checks

Each time the system performs diagnostic tests or an instrument health check (IHC), it creates a
report of the results.
1 From the Home screen, press Settings.
The
2 Press Self Diagnostic.
The
3 Press Browse Report.
The
Magnis and MagnisDx NGS Prep System User Guide
Settings screen
opens.
Instrument Diagnostic screen
Diagnostic Report Explorer screen
opens.
opens.