Product Inspection; Safety Precautions - Agilent Technologies Magnis G9710A User Manual

Ngs prep system, for in vitro diagnostic use, for research use only, not for use in diagnostic procedures
Table of Contents

Advertisement

Product Inspection

Upon receipt of the Magnis/MagnisDx NGS Prep System, carefully inspect the product box for any
visible signs of damage. If damage to the product box is detected, contact
Technical
Allow the shipping container for the Magnis/MagnisDx NGS Prep System to come to room
temperature prior to unpacking.

Safety Precautions

The Magnis/MagnisDx NGS Prep System is designed for safe operation when it is used in the
intended manner. Using the system for purposes other than those intended may impair those
safety protections.
Installation
• The system must be installed by an Agilent engineer or Agilent authorized service provider.
• Do not attempt to manually lift the instrument. To move the instrument, use an automated
forklift or lift table that can lift at least 100 kg.
• When adjusting the placement of the instrument on the laboratory bench, take care to avoid
pinching hazards on the instrument.
• Retain the locking jigs that secure the gantry during shipment. Anytime the instrument is
moved, secure the gantry using the jigs.
Electrical
Standard electrical safety precautions should be applied, including the following:
• Per North American and IEC requirements, install the instrument in a location where branch
circuit protection is available in the power network.
• Install the instrument with the Agilent supplied power cord that is compatible with the
electrical outlets in your region. Do not substitute with a power cord from another source.
• Install the instrument in a location where the power cord can be easily reached for quick
disconnection of the power supply.
• Install the instrument away from any flammable sources.
• Make sure that the ventilation slots of the instrument are free to vent at all times. Maintain
10 cm of cleared space on each side of the instrument and 18 cm of cleared space at the back
of the instrument.
• Connect the power cord to a wall output that provides 100–240 VAC, 1000 W.
• Make sure that proper voltage is supplied before turning on the instrument for the first time.
• Connect the instrument to a grounded socket. Do not operate the instrument from a power
outlet that has no ground connection.
• Do not connect the instrument to the same circuit as other devices with a high current draw
(e.g., freezers, centrifuges). If possible, connect the instrument to an independent or dedicated
AC circuit.
• Do not touch any switches or outlets with wet hands.
Magnis and MagnisDx NGS Prep System User Guide
Support.
Agilent Worldwide
12