Use a wiping pattern that will cover all exposed surfaces. Verify that the wipe contains
sufficient cleaning agent to wet the surfaces and if not, obtain a fresh wipe or reapply
additional cleaning agent to the wipe. If a wipe becomes visibly soiled, change to a clean wipe.
a Instrument interior: Open the instrument door and wipe down the areas of the deck that are
Figure 32 Deck surfaces to be wiped down
b Instrument Exterior - Front: Close the instrument door. Using a laboratory wipe moistened
Magnis and MagnisDx NGS Prep System User Guide
outlined in yellow in
Figure 32
areas again using a laboratory wipe moistened with 70% isopropyl alcohol.
with 70% isopropyl alcohol, gently wipe down the surfaces that are outlined in yellow and
red in
Figure
33. These areas include the touchscreen, power button, waste bin door
(including depression used to assist in opening/removing), and the front plexiglass. Pull out
the waste bin drawer and wipe down the interior of the waste bin. Wipe down the area
around the USB ports (do not wet the electrical contact points). All areas outlined in yellow
require wiping down. Areas outlined in red require particular attention for careful cleaning.
using a diluted bleach wipe. Then, wipe down these same
53