Hardware Usage Tracking Screen; Instrument Diagnostic Screen - Agilent Technologies Magnis G9710A User Manual

Ngs prep system, for in vitro diagnostic use, for research use only, not for use in diagnostic procedures
Table of Contents

Advertisement

Hardware Usage Tracking screen

Purpose: To track the usage of the system's UV tube. For Agilent engineers and Agilent authorized
service providers, to reset the tracker following replacement of the UV tube. See instructions in
"Replacing the UV Tube and Viewing UV Tube Usage"
To open: From the Home screen, press Settings. On the Settings screen press Hardware Usage
Tracking.
Figure 51 Hardware Usage Tracking screen
UV Tube
Next to this check box is displayed the number of hours, minutes, and seconds of usage for the
UV tube. Marking the check box enables the Reset button
Reset
This button resets the usage tracker for the UV tube to zero. Once reset, the usage is displayed as
0h 00m 00s. This action should only be performed by an Agilent engineer or Agilent authorized
service provider following replacement of the UV tube.
NOT E
Close
This button returns you to the Settings screen.

Instrument Diagnostic screen

Purpose: To run diagnostic tests on the instrument and access reports for previous diagnostic
tests and instrument health checks. See instructions in
page 39.
To open: From the Home screen, press Settings. On the Settings screen press Self Diagnostic.
Magnis and MagnisDx NGS Prep System User Guide
The lifespan of the UV tube is 630 hours. After 630 hours of usage, when a
decontamination cycle is initiated, the system notifies you to replace the UV
tube.
on page 56.
"Perform instrument diagnostic tests"
on
73