7 When prompted on the touchscreen for a tip box and tip, open the instrument door, add a new
tip box (with lid off) to the indicated platform, and add a tip to the micropipettor at the indicated
barrel position. Close the door, then press Next to continue.
8 When prompted on the touchscreen to remove the tip box, open the instrument door and
remove the tip box. Close the door, then press Next to continue.
The auto-teaching process continues until completion.
Include teach-point verification in the IHC
The system can include teach-point verification as part of the initial instrument health check (IHC)
that is performed after powering up the instrument. If the system detects an error during
teach-point verification, the corrective action is to run auto-teaching.
1 From the Home screen, press Settings.
The
2 Press System Settings.
The
3 Press Other Settings.
The
Figure 27 Other Settings screen
4 Mark the check box labeled Verify teach points during Instrument Health Check.
5 Press Close to save your changes.
Magnis and MagnisDx NGS Prep System User Guide
Settings screen
opens.
System Settings screen
Other Settings screen
opens.
opens.
45