To Install the Optional FID Vent Chimney
WA RNING
C A U T I O N
Maintaining Your GC
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Use if methylene chloride solvent or other sample produces
corrosive byproducts.
1
Remove the detector cover.
2
Insert the O-ring into bottom of FID vent chimney.
3
Insert the long end of the FID vent chimney up through the
bottom of the detector cover so that the holes in the vent
chimney align with the holes in the cover.
4
Secure the vent chimney to the cover using the two
self-tapping screws.
5
Maintaining the FID
Detector cover
Vent chimney
Screw
O-Ring