Export Files Screen - Agilent Technologies Magnis G9710A User Manual

Ngs prep system, for in vitro diagnostic use, for research use only, not for use in diagnostic procedures
Table of Contents

Advertisement

Table 9
Button
Firmware Update
Other Settings

Export Files screen

Purpose: To export post run data files, system logs, error logs, and debug logs.
To open: From the Home screen, press Settings. On the Settings screen press Export.
Figure 45 Export Files screen
File folders
The panel on the left of the screen lists the folders from which you can export files. The folders are
described below.
• Post run data folder: Contains data files for completed runs.
In the RunData folder, subfolders are organized by user account, then by year, then numerically
by month. For file export, select the desired subfolder; the system will export all post run data
files in that folder.
• System Log folder: Contains log files of system-level actions, such as powering on of the
instrument and diagnostic tests.
• Error Log folder: Contains log files of system errors.
• Debug Log folder: Contains log files of debugging actions. The files are organized by
instrument module into separate subfolders.
• Raw Data Log folder: Contains log files about the controller area network (CAN)
communication messages received and sent by the touchscreen application.
• Self Test Log folder: Contains log files of the CAN communication messages and results of the
most recent self diagnostic tests.
Magnis and MagnisDx NGS Prep System User Guide
Settings screen buttons (continued)
Description
Opens the
firmware software from a connected USB disk drive or by connecting to the
Agilent website.
Opens the
during the IHC.
"Firmware Update screen"
"Other Settings screen"
which has a setting to verify teach points
for installing a new version of the