Figure 22 Diagnostic Report Explorer screen
4 On the left side of the screen, select the folder containing the report type of interest.
The Diagnostic Report folder contains reports from instrument diagnostic tests. The IHC
Report folder contains reports from instrument health checks.
The right side of the screen displays the reports in the selected folder.
5 On the right side of the screen, locate the report you want to view. Press the report to select it.
6 Press View.
The screen displays the report in the
contains summary information, including the date and time of the tests. The table lists the
tests that were performed, organized by system component, and the result of each test
(Passed, Failed, or Skipped). Note that three of the quick ramping tests are skipped by default.
Figure 23 Diagnostic Test Report screen
7 Press Close to close the report.
You are returned to the Diagnostic Report Explorer screen.
Users with Advanced access can delete older reports from the Diagnostic Report Explorer screen.
Locate the report you want to delete, then press the report to select it. Press Delete to delete the
selected report. Note that you cannot delete the most recent report.
Magnis and MagnisDx NGS Prep System User Guide
Diagnostic Test Report