Environmental Requirements; Installation Requirements - Agilent Technologies Magnis G9710A User Manual

Ngs prep system, for in vitro diagnostic use, for research use only, not for use in diagnostic procedures
Table of Contents

Advertisement

Electrostatic Discharge
The Magnis/MagnisDx NGS Prep System is static sensitive. Electrostatic discharges greater than
8000 volts may interfere with the normal operation of the USB ports on the instrument. Handling
precautions are required when working in high static environments. Wear a grounded wrist strap
and take other anti-static precautions prior to making contact with the instrument in high static
environments. ESD STM5.1-1998 Class 3B.

Environmental Requirements

Operating temperature
Keep the ambient temperature between 15°C and 25°C.
Operating humidity
Keep the humidity levels between 30% and 70% non-condensing.

Installation Requirements

• Place the instrument in a post-PCR laboratory environment.
• Do not place the instrument in proximity to other lab equipment that is sensitive to vibrations
or that generates vibrations during use.
• Maintain the following minimum distances for cleared space around the instrument.
• Install the instrument with the Agilent supplied power cord. Do not substitute with a power
cord from another source.
• Install the instrument in a location where the power cord can be easily reached for quick
disconnection of the power supply.
• Install the instrument away from any flammable sources.
Review the
safe operation of the instrument.
NOT E
Magnis and MagnisDx NGS Prep System User Guide
Instrument sides: 10 cm (4 inches) on each side to allow for proper ventilation from the side
vents.
Back of the instrument: 18 cm (7 inches) at the back to allow for proper ventilation from the
back vent.
Front of the instrument: 5 cm (2 inches) in front to avoid unintentional contact with the
power switch.
Top of the instrument: 111 cm (44 inches) above to allow for opening the door.
"Electrical"
safety precautions on
After installation is complete, avoid moving the instrument or making any
adjustments to the placement of the instrument as it may disturb some of the
settings made by the Agilent engineer or service provider.