Pentax EPK-i7010 Instructions For Use Manual page 92

Table of Contents

Advertisement

Symptom
The endoscope buttons do not
work.
Cannot release freeze.
The time stamp of the saved
still image file/video image file
does not match the time when
the image was captured.
Seek bar or seek time of video
image file is not displayed
appropriately by reproduction
software.
Image processing is not
reflected in the left side screen
after the Twin mode setup.
It cannot be copied from
Endonet during a "VCU ready."
7
display.
Image is deteriorated when the
saved recording file is played
with PC.
"No Signal" is displayed in a
still image.
*1: For details, refer to the Instructions for Use for the monitor.
*2: For details, refer to the Instructions for Use for the endoscope.
NOTE
 If there is no response even if you press any key on the keyboard, turn off the processor and contact a PENTAX
Medical service facility.
 If the problem persists even after taking the measures described in the corresponding solutions above, contact a
PENTAX Medical service facility.
87
Troubleshooting
Possible Cause
Cannot communicate properly with the
endoscope.
System malfunction caused by static
generation.
This processor's "Time zone setting
(P.59)" or "Date & Time (P.58)" differs
from PC settings.
As time property of video image file
can't be read out appropriately by
reproduction software.
Setup of Twin mode was made after
freezing.
RS-232C cable is not connected.
This is generated originating in a
compression codec processing while
setting up the compression ratio Best
quality.
Checked [Add external.] box of [Option
(Still image)] while a still image was
captured in a state of no inputting an
external input image in P-in-P mode.
Possible Solution
Press [Scope eject] button and remove
the endoscope from the processor by
aligning the endoscope locking lever to
the OPEN position, and then connect
the endoscope again. If the function is
still not available, release the
endoscope locking lever and the
endoscope's distal tip should be
straightened to its neutral position.
Then, carefully withdraw the
endoscope from the patient.
Release the endoscope locking lever
and the endoscope's distal tip should
be straightened to its neutral position.
Then, carefully withdraw the
endoscope from the patient.
Set up this processor's "Time zone
setting (P.59)" and "Date & Time
(P.58)" to become the same as PC
settings.
When replayed the still image file/the
video image file captured by PC having
been set a different ""Date & Time" or
"Time zone setting", the time stamp on
the still image file/video image file may
be different from the date and time
captured. Therefore, confirm the date
and time displayed on the PC monitor.
Use other reproduction software to
replay.
Release freezing and check Twin
mode being set, then execute freezing
again.
Connect RS-232C cable correctly and
try to copy again.
Use setting it up with the compression
rate for High quality or Standard
quality.
Capture a still image again after check
that an external input image is inputted
correctly.
Or when an external input is not used,
cancel P-in-P mode and capture a still
image again.
Reference
Page
P.66
P.58
P.59
P.65
P.41
P.64