NOTE
 For peripheral devices, use the recommended devices shown above or equivalent products.
 PENTAX Medical recommends the devices shown above based on the independent test results.
 The processor or the peripheral devices may not operate normally by the combination of their connections including
the recommended devices.
 When any connected peripheral devices are added, use the processor and the peripheral devices after confirming
their normal operations by checking their connected combinations.
 The data may not be recorded properly if non-compatible external hard drive or non-compatible USB flash memory
is used, or if the USB flash memory is connected to the USB port of this product using USB extension cable.

Installation

 Installing the Processor in a Cart
Install the processor and peripheral devices in a cart while referring to the installation examples in "System Configuration" (P.9).
Place the cart on a stable and level surface.
WARNING
 The processor is an electro-medical device incorporating precision components and sophisticated circuitry so do not
install it in a location where it will be subject to harsh conditions, including excessive vibrations and/or severe impact.
Never drop the processor or subject it to a severe impact as doing so could compromise the functionality and/or safety
of the unit. Should this equipment be mishandled or dropped, do not use it. Return it to an authorized PENTAX
Medical service facility for inspection and repair.
 Because of the adverse effect that one equipment could cause to another equipment, it is possible that the correct
operation of this processor could be compromised if another equipment is placed right nearby or stacked on top of
the processor. If such situations cannot be avoided for extenuating reasons, be sure to check the correct operation
of each equipment before using them in clinical use.
 Do NOT block the ventilation grids on the sides and at the back of the processor. In particular make sure to separate
the ventilation grids on the processor's left side at least 15 cm (6 inches) from the wall. Blocking ventilation may cause
equipment temperature to rise.
 When you raise this processor, hold the portion from the both sides shown in the figure.
 Make sure to firmly connect each connector to the specified position. Failing to do so could cause certain function to
be unavailable.
 Install the processor on a stable and level surface (cart, counter, stand, etc.).
- Avoid places where the processor may be splashed with liquid.
- Never use the processor where explosive or flammable gases are present.
- Do not install the processor where it will be exposed to high temperature and humidity or direct sunlight.
- Do NOT use this equipment by installing it to stand vertically. Doing so will hinder the air to vent through the
ventilation grid, and eventually it will become the cause of product damage.
 Never install, operate, or store electro-medical equipment in a dusty place. The accumulation of dust within the
equipment may cause a malfunction, smoke, or ignition.
 Do not block the air intake vents on the back and the sides of the processor.
 When moving the processor, do not hold the endoscope locking lever.
 Connect a water bottle assembly (OS-H5) before using the processor.
 When loading the processor on a shelf of the cart, provide sufficient space for attaching or detaching the water bottle
assembly (OS-H5) easily.