Disassembly Guide
48
7. Install the display PCB and CPU PCB into the NPB-40 case. Ensure that the
battery springs slip into place. See Figure 11.

Figure 11: Battery Spring Alignment

Note: The longer screws install at the top of the case. See Figure 12, items 1
and 4.
8. Align the front and rear cases. See Figure 12.

Figure 12: Installing The Case Screws

9. Install the four screws (Figure 12, items 1 through 4).