Performance Verification
Test #1: BPM
26
•
Pulse Amplitude indicator - show low level modulation (low amplitude
pulse amplitude indicator)
1. Press the SRC-MAX % PULSE RATE selection button. The SRC-MAX
PULSE RATE 200 LED lights.
2. The NPB-40 bpm increases to 200 and stabilize at 200 bpm. The test pass
criteria is 197 to 203 bpm inclusive.
3. The NPB-40:
•
displays 75 %SpO
•
displays 200 bpm (pass criteria is 197 to 203 bpm inclusive)
•
alarms
•
Pulse Amplitude indicator displays low level modulation
4. Press the SRC-MAX PULSE RATE select button. The SRC-MAX PULSE
RATE 60 LED lights.