Disassembly Guide
50
4. Connect the new front case assembly ribbon cable to the CPU PCB connector
J3. See Figure 14.

Figure 14: Front Case Ribbon Cable Connection

5. Lock the ribbon cable connector. See Figure 15.

Figure 15: Locking CPU PCB Connection to J3

1 — J3
1 — CPU PCB
2 — J3 lock
3 — Lock direction
4 — Unlock direction