NPB-40
Caution: A cable runs between the NPB-40 front and rear case assemblies.
Use care when separating the cases. Do not apply stress to the cable.
3. Separate the NPB-40 front and rear case assemblies. See Figure 6.
Caution: Failure to unlock connector J3 on the CPU PCB before attempting
to remove the front case flex circuit could damage the flex circuit.
4. Unlock connector J3 on the CPU PCB. See Figure 6 and Figure 7.

Figure 7: CPU PCB Connector J3 Lock

5. Disconnect the ribbon cable from CPU PCB connector J3.
1 — Connector J3

Figure 6: Cases Separated

1 — CPU PCB
2 — J3 lock
3 — Lock direction
4 — Unlock direction
Disassembly Guide