Test #2: SpO
NPB-40
5. The NPB-40 pulse rate decreases to 60 and stabilize at 60 bpm. The test pass
criteria is 57 to 63 bpm inclusive.
6. The NPB-40:
•
displays 75 %SpO
•
displays 60 bpm (test pass criteria is 57 to 63 bpm inclusive)
•
alarms
•
Pulse Amplitude indicator displays low level modulation
2
1. Press the SRC-MAX %SpO2 select button. The SRC-MAX %SpO2 90
LED lights.
2. The NPB-40 displays three dashes until the SRC-MAX stabilizes at 90
%SpO
. The test pass criteria is 88 to 92 %SpO
2
3. The NPB-40:
•
displays 90 %SpO
•
displays 60 bpm
•
alarms
•
Pulse Amplitude indicator displays low level modulation
4. Press the SRC-MAX %SpO2 select button. The SRC-MAX %SpO2 75
LED lights.
2
(pass criteria is 88 to 92 %SpO
2
Performance Verification
inclusive.
2
inclusive)
2
27