Specifications
66
Audible Indicators (Continued)
Audible
Parameter
Indicator
High Priority
Volume level
Alarm
Pitch (±30 Hz)
On pulse width
(±20 msec)
Off Interval (±20 msec)
Number of pulses in burst 1
Repetition Pause
(±2 sec.)
Priority
Medium Priority
Volume level
Alarm
Pitch (±30 Hz)
On pulse width
(±20 msec)
Off Interval (±20 msec)
Number of pulses in burst 1
Repetition Pause
(±2 sec.)
Priority
Low priority
Volume level
Alarm
Pitch (±30 Hz)
On pulse width
(±20 msec)
Off Interval (±20 msec)
Number of pulses in burst 1
Repetition Pause
(±2 sec.)
Priority
Value
Adjustable, 42 to 52
dB(A), at one meter
1200 Hz
250 msec
80 msec
0 sec.
9
Adjustable, 42 to 52
dB(A), at one meter
752 Hz
400 msec
300 msec
0 sec.
10
Adjustable, 42 to 52
dB(A), at one meter
500 Hz
400 msec
3200 msec
0 sec.
11