3.2 Installing the Converter

Converter Housing
CAUTION
IM 11M10A01-03E
(1) Installation Site
The converter should be installed in places where:
• Easy access to display and key operations can be achieved.
• Inspection and maintenance can be easily performed.
• Ambient temperature does not exceed 50 C and temperature fluctuations are minimal:
Avoid exposure to direct sunlight or radiant heat.
• Humidity is moderate (40 to 75% RH recommended), and no corrosive gases or metal
dust are present.
• Mechanical vibration should be negligible.
• No electromagnetic field should be present.
• No exposure to rain or water is allowed.
(2) Mounting
The converter is panel mounted (copperplate of 1 to 10 mm thickness) using the
supplied brackets.
Procedure
• Insert the unit through the mounting hole from the front.
• Mount the converter on the panel using brackets. The brackets should be fixed to the
designated position on the right and left sides of the housing by tightening the screws.
See Figure 3.2.
• The converter should be upright and horizontal.
• Ensure that the converter bracket screws are tightly secured so there is no movement.
Figure 3.2 Converter Mounting
Over-tightening the screws may distort the case or damage the brackets.
Panel
Removal of Protective Film
Upon completing installation, remove the
Mounting Brackets (2 sets)
transparent film from the operation panel of
This film is for protecting the panel from
damage during shipment.
3. Installation
F302.EPS
3-3