YOKOGAWA vigilantplant OX102 User Manual page 54

Current limit type oxygen analyzer
Table of Contents

Advertisement

[10000 ppm O
Calibration]
2
( ) CAL4 is for 0-10000 ppm
*
range calibration. Assuming
the O
concentration of
2
calibration gas is 9310 ppm,
enter this value using
and
keys.
6-20
Continued from CAL.3
Indication
%
ppm
CAL.4
CAL
%
ppm
Entering calibration value
Calibration
completed
%
ppm
CAL5 (to the following page)
Stop supplying calibration gas No.5
Supplying calibration gas No.6 (O
allowance: 8500-9500 ppm)
ppm+CAL
ENT
Calibration value
CAL
ENT
*1
*2
Calculating
YES
ERROR
NO
Calibration
accepted
9310
*1 Entered to skip calibration
*2 Entered to enter calibration value
*3 Entered to return to measurement mode
Measurement Mode
concentration
2
ppm+CAL
*3
Changing
value
IM 11M10A01-03E