YOKOGAWA vigilantplant OX102 User Manual page 53

Current limit type oxygen analyzer
Table of Contents

Advertisement

[1000 ppm O
Calibration]
2
( ) CAL3 is for 0-1000 ppm
*
range calibration.
Assuming the O
concentration
2
of calibration gas is 995 ppm,
enter this value using
and
keys.
IM 11M10A01-03E
Continued from CAL.2
Indication
%
CAL.3
ppm
CAL
%
ppm
Entering calibration value
Calibration completed
%
ppm
CAL4 (to the following page)
Stop supplying calibration gas No.3
Supplying calibration gas No.5 (O
concentration allowance: 900-1100 ppm)
ENT
ppm+CAL
Calibration
value
CAL
ENT
*1
*2
Calculating
Changing
value
YES
ERROR
NO
Calibration
accepted
995
*1 Entered to skip calibration
*2 Entered to enter calibration value
*3 Entered to return to measurement mode
6. Operation
Measurement Mode
ppm+CAL
*3
2
F6314.EPS
6-19