5.

Component Functions

5.1

Sensor

IM 11M10A01-03E
This chapter describes functions of main equipment of the OX102 oxygen analyzer.
The sensor uses a zirconia current limiting system, enabling direct insertion of the sensor
at elevated temperatures up to 150 C. 0 to 250 C within 50 mm of sensor tip. The sensor
should be used at a measuring point pressure of 1013
sensor, enter the data in the Sensor Parameter Table (attached to the sensor control
block) into the converter. This ensures sensor compatibility.
Figure 5.1 Sensor
* Sensor Parameter Table (Example)
Sensor Parameter Table
Sensor No. YA-0007
Gain 1
1.015
Gain 2
A
1.043
B
-65
A
1.002
Gain 3
B
-25
Gain 4
A
0.990
B
-19
Gain 5
A
1.429
B
-17
When replacing the sensor, enter values in the
Sensor Parameter Table into the converter.
See Section 6.3 (1) "Setting Sensor Parameter"
for procedure.
5. Component Functions
40 hPa abs. When replacing the
Sensor Parameter Table *
F501.EPS
T5101.EPS
5-1