YOKOGAWA vigilantplant OX102 User Manual page 44

Current limit type oxygen analyzer
Table of Contents

Advertisement

6-10
%
*
ppm
%
ppm
*
Gain 6 (Gln. 6) is not user configurable.
(2) Automatic/Manual Switching
In this mode, either automatic or manual range switching can be selected. Set either
mode following the procedure shown in the flow chart below.
When the autoswitching is selected, the mode of "Configuring the 4-20 mA output
signal" described in Section (3) is automatically determined.
Indication
%
ppm
%
ppm
ENT
Mode
accepted
( ) Use [ ] and [ ] keys to change the mode of AUTO or SELC.
*
AUTO : Automatic range mode
SELC : Fixed range mode
Figure 6.5 Flow Chart of Automatic/Manual Switching Setting
Continued
GIN.6
ppm+CAL
ENT
%
All GAIN Value
%
Operational Flow Chart
Measured
Value
+ 4-20mA
( )
*
AUTO or SELC
ppm+CAL
Changing
Mode
To measurement mode
ppm+CAL
F6307E.EPS
IM 11M10A01-03E
F6306E.EPS