YOKOGAWA vigilantplant OX102 User Manual page 46

Current limit type oxygen analyzer
Table of Contents

Advertisement

6-12
(4) Configuring the 1-5 V output signal
This mode is to configure the 1-5 V output signal to correspond to a fixed range.
Configure the signals following the procedure shown in the flow chart below.
This output signal is always set to correspond to a fixed range. Therefore, even if
automatic range switching is selected, this output signal should be configured for each
range.
Indication
%
ppm
%
ppm
( * ) Use [ ] and [ ] keys to configure Out.1 to 4.
Out.1 : Full scale 25% O
Out 2 : Full scale 10000 ppm O
Out 3 : Full scale 1000 ppm O
Out 4 : Full scale 100 ppm O
Figure 6.7 Flow Chart of 1-5 V Output Signal Setting
Operational Flow Chart
Measured
Value
+ 4 - 20mA
Out . 1 - 4
( * )
4-20mA
ENT
Output
Changing
accepted
value
2
2
2
2
ppm+CAL
( * )
F6309E.EPS
IM 11M10A01-03E