Appendix 2. Calibration Gas Kits

K9424JA, K9424JB, K9424JR, K9424JS Calibration Gas
Kits for OX100/102

1. Connecting the needle valve assembly to the gas cylinder

IM 11M10A01-03E
The calibration gas kits are designed for calibration of the OX100 Series Oxygen
Analyzers.
The kit comprises a calibration gas (filled pressure of 0.7 MPa, 6 one-litter cylinders)
and a needle valve assembly. The calibration gas is available in two types: 900 to 1100
ppm O
and 8500 to 9500 ppm O
2
application.
Take the following steps to connect the gas cylinder to the needle valve assembly.
(1) Turn the handle of the needle valve counterclockwise to pull the needle tip inside the
gasket to the position shown in Figure 1.
(2) Loosen the locknut of the needle valve. Screw the valve into the mouth of the gas
cylinder in a clockwise direction with fingers until it meets resistance and then screw
the locknut down (See Figure 2).
Figure 1
Needle
Packin
Container
Figure 2
(N
based). Prepare either one according to your
2
2
Valve
Lock-nut
AP201E-1.eps
Appendix
A-5