5.3 Sampling Unit (Optional)

Sample Inlet, Rc1/4
Disk filter : K9424GQ
Sample Outlet, Rc1/4
Activated
Carbon Filter
IM 11M10A01-03E
The sampling unit comprises membrane filters, suction pump, flowmeter and activated
carbon filter to absorb organic solvents. When a sample gas contains organic solvents,
use this sampling unit to filter out such solvents. The expected life of the activated
carbon filter is approximately 50 hours when a sample gas contains 500 ppm of isopro-
pyl alcohol and flows at 500 ml/min.
Used eight hours a day, suction pump life will be approximately one year.
Flowmeter
Sensor Connector
(for 8 mm O.D.)
Switching Power Supply
Disk Filter
(K9424GQ)
Sample Gas
Inlet
Pump
Sample Gas
Outlet
: Rc 1/4
: Provided by customer
(Stop valves must be installed at the sample gas inlet and outlet.)
Figure 5.3 Sampling Unit
Power Connector
Sampling Conditions
Pressure at sampling point
Sample gas flow rate
Pressure at sample outlet
Gas temperature
Organic solvent content
Dust content
Power supply
Piping connection
Needle Valve
OX102 Sensor
Switch
Felt Filter
Flowmeter,
500 ml/min
(K9346WH)
Activated Carbon Filter
Needle Valve
5. Component Functions
: 1013
40 hPa abs
: 500
100 ml/min
: 1013
40 hPa abs
: 50 C max.
: ppm to a few percentage
3
: 5 mg/Nm max.
: 100 to 240 V, 50/60 Hz
: Rc1/4
Switch
Power Supply
Sensor
Insertion Port
F5302.eps
F5301.EPS
5-3