(3) Configuring the 4-20 mA output signal
This mode is to configure the 4-20 mA output signal for the fixed range. Configure the
signals following the procedure shown in the flow chart below.
This mode is automatically determined if the automatic/manual range switching mode is
set to autorange mode.
Indication
%
ppm
%
Out . 1 - 4
ppm
4-20mA
When AUTO is selected, the measurement mode remains unchanged.
(*1)
Use [ ] and [ ] keys to configure Out.1 to 4.
(*2)
Out.1 : Full scale 25% O
Out 2 : Full scale 10000 ppm O
Out 3 : Full scale 1000 ppm O
Out 4 : Full scale 100 ppm O
Figure 6.6 Flow Chart of 4-20 mA Output Signal Setting
Operational Flow Chart
Measured
Value
4-20mA
(*1)
(*2)
ENT
Changing
Output
value
accepted
2
2
2
2
6. Operation
ppm+CAL
(*2)
F6308E.EPS
6-11