YOKOGAWA vigilantplant OX102 User Manual page 57

Current limit type oxygen analyzer
Table of Contents

Advertisement

[500 ppm calibration] Indication
[1000 ppm calibration]
IM 11M10A01-03E
%
ppm
ENT
Entering calibration
value
%
Calibration value
ppm
ENT
Measured
%
value
ppm
ENT
Calculating
Calibration
ERROR
completed
%
ppm
%
1000
ppm
ENT
%
Calibration value
ppm
ENT
%
Measured
value
ppm
ENT
Calculating
ERROR
Calibration
completed
%
ppm
10000
%
ppm
Stop supplying calibration gas No.3 and
500
supply No.4
(O
concentration allowance: 400-600 ppm)
2
ppm+CAL
ppm + CAL
Changing
value
YES
NO
Waiting for 1000 ppm calibration
Stop supplying calibration gas No.4 and
supply No.5
ppm+CAL
(O
concentration allowance: 900-1100 ppm)
2
ppm + CAL
Changing
value
YES
NO
Waiting for 10000 ppm calibration
Note: To return to measurement mode, press [ppm + CAL]
6. Operation
Measurement mode
F6318E.EPS
6-23

Advertisement

Table of Contents
loading

Table of Contents