( ) If the entered high alarm setting is less than the low alarm setting, the display will show
*
and the entered setting will flash to prompt reentry. An appropriate value should be reentered.
6-14
Continued
%
Low alarm
ppm
ALM
%
ppm
(6) Resetting the sensor parameters to the default values
This is to reset the sensor parameters (A and B values for each gain, see Section 5) to
the factory default values.
Indication
%
Measured
Value
ppm
%
FSET
ppm
%
Figure 6.9 Flow Chart of Sensor Parameter Resetting
ppm+CAL
ENT
Low alarm setting
ENT
ALM
No
High Alarm
>
=
Low Alarm
Yes
Value
accepted
Operational Flow Chart
+ ENT
+
ppm+CAL
ENT
To the factory
default values
F6321E.EPS
Measurement mode
ppm
%
ppm+CAL
Changing
value
E- - -
for about 3 seconds
F6311E.EPS