Appendix 1. Sampling Unit; K9424Ga Sampling Unit For Ox100/102; Specifications - YOKOGAWA vigilantplant OX102 User Manual

Current limit type oxygen analyzer
Table of Contents

Advertisement

Appendix 1. Sampling Unit

K9424GA Sampling Unit for OX100/102

Notes

1. Specifications

IM 11M10A01-03E
This sampling unit is designed for the OX100 Series Oxygen Analyzers when they are
used where a sample gas contains a small amount of organic solvents (isopropyl alco-
hol), such as in reflow furnaces.
The sampling unit comprises two membrane filters to eliminate dust and mist, a suction
pump to draw a sample gas, an activated carbon filter to eliminate organic solvents, and
a flowmeter to monitor sample gas flow rates.
If repair is required, the is equipment will need to be returned to the factory. For repair
or maintenance, contact your Yokogawa representative.
Pressure at sampling point
Pressure at sample outlet
Gas temperature
Organic solvent content
Dust content
Life expectancy of activated carbon: The expected life of the activated carbon filter is
Power supply
Piping connection
Operating environment temperature range: 0 to 40 C
Life of disk filter
Pump life
Included Accessories
1. Power cord (1)
A1007WD (AC 100 V standard)
A1009WD (AC 200 V by special order)
2. Fuse (2)
A1109EF (rating 1A)
3. Instruction Manual (1)
K9424JY
: Atmospheric pressure
: Atmospheric pressure
: 50 C max.
: ppm to a few percentage level of isopropyl alcohol
(IPA) equivalent
3
: 5 mg/Nm
max.
approx. 50 hours when a sample gas contains 500
ppm of isopropyl alcohol and flows at 500 ml/min.
: 100 to 240 V AC, 50/60 Hz
: Rc1/4
: When specified flow is no longer possible. Replace
as early as possible.
: Approximately one year when used eight hours a
day.
Appendix
A-1