YOKOGAWA vigilantplant OX102 User Manual page 58

Current limit type oxygen analyzer
Table of Contents

Advertisement

[10000 ppm O
Calibration]
2
6-24
Indication
%
ppm
Entering calibration
value
%
ppm
%
ppm
Calibration
completed
%
ppm
(2) Moving zero point
This is to translate the output function: to change value b of the characteristic function, y
= ax + b. The range for moving the zero point is from -999 to +999 ppm.
Indication
Measured
%
value
ppm
%
Increment/decrement
ppm
Calculating/
Configuring
Note: To return to measurement mode, press [ppm + CAL]
Figure 6.13 Flow Chart of Moving Zero Point
Stop supplying calibration gas No.5
and supply No.6
10000
(O
ppm+CAL
ENT
Calibration value
ENT
Measured
value
ENT
Changing
Calculating
value
ERROR
YES
NO
Note: To return to measurement mode, press [ppm + CAL]
Operational Flow Chart
Changing
value
concentration allowance: 8500-9500 ppm)
2
Measurement mode
ppm + CAL
F6319E.EPS
F6320E.EPS