YOKOGAWA vigilantplant OX102 User Manual page 52

Current limit type oxygen analyzer
Table of Contents

Advertisement

[100 ppm O
Calibration]
2
(*) CAL2 is for 0-100 ppm
range calibration.
Assuming the O
concentration of
calibration gas is 110 ppm,
enter this value using
and
6-18
Indication
%
ppm
2
keys.
%
ppm
Entering calibrated value
Calibration
accepted
%
ppm
CAL3 (to the following page)
Continued from CAL.1
Stop supplying calibration gas No.1
CAL.2
ppm+CAL
CAL
ENT
Calibrated
value
CAL
ENT
Calculating
ERROR
NO
Calibration
completed
110
*1 Entered to skip calibration
*2 Entered to enter calibration value
*3 Entered to return to measurement mode
Measurement Mode
Supplying calibration gas No.3
(O
concentration in the range: 80-120 ppm)
2
ppm+CAL
Changing
value
YES
F6313E.EPS